AC192 Series

Compact Multipurpose Accelerometer, Top Exit 2 Pin Connector, 100 mV/g, ±10%
AC192-1D
2 Pin Connector
Connector Pin Polarity
A (+) Signal/Power
B (-) Common
Stock Product
AC192-2D
CB103 Integral Cable
Conductor Polarity
Red (+) Signal/Power
Black (-) Common
Shield Cable Drain Wire
Built To Order
AC192-3D
CB206 Armored Integral Cable
Conductor Polarity
Red (+) Signal/Power
Black (-) Common
Shield Cable Drain Wire
Built To Order
AC192-6D
CB611 Heavy Duty Armored Integral Cable
Conductor Polarity
Red (+) Signal/Power
Black (-) Common
Shield Cable Drain Wire
Built To Order
Specifications Standard Metric
Part Number AC192 M/AC192
Sensitivity (±10%) 100 mV/g
Frequency Response (±3dB) 24-780,000 CPM 0,4-13000 Hz
Frequency Response (±10%) 60-540,000 CPM 1,0-9000 Hz
Frequency Response (±5%) 600-300,000 CPM 10-5000 Hz
Dynamic Range ± 80 g, peak
Electrical
Settling Time <2 Seconds
Voltage Source (IEPE) 18-30 VDC
Constant Current Excitation 2-10 mA
Spectral Noise @ 10 Hz 8 µg/√Hz
Spectral Noise @ 100 Hz 4 µg/√Hz
Spectral Noise @ 1000 Hz 2 µg/√Hz
Output Impedance <100 ohm
Bias Output Voltage 10-14 VDC
Case Isolation >108 ohm
Specifications Standard Metric
Environmental
Temperature Range -58 to 250°F -50 to 121°C
Maximum Shock Protection 5000 g, Peak
Electromagnetic Sensitivity CE
Sealing Welded, Hermetic
Submersible Depth 200 ft. 60 m
Physical
Sensing Element PZT Ceramic
Sensing Structure Shear Mode
Weight 1.8 oz 51 grams
Case Material 316L Stainless Steel
Mounting 1/4-28
Connector (Non-Integral) 2 Pin MIL-C-5015
Resonant Frequency 1,560,000 CPM 26000 Hz
Mounting Torque 2 to 5 ft. lbs. 2,7 to 6,8 Nm
Mounting Hardware 1/4-28 Stud M6x1 Adapter Stud
Calibration Certificate CA10

Typical Frequency Response

Frequency Response Chart